Measurement of HDL-cholesterol in serum by a new enzymatic method.

Loading...
Thumbnail Image
Date
2001-07
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
High density lipoprotein (HDL) are responsible for the reverse transport of cholesterol from the peripheral cells to the liver. Here, cholesterol is metabolised. Monitoring of HDL-C in serum is of clinical importance since an inverse correlation exists between serum HDL-C concentrations and the risk of atheroselerotic disease. Elevated HDL-C concentrations are protective against coronary heart disease. In this study, an automated enzymatic method for the direct determination of HDL-C in serum was found to have good precision and accuracy. Reproducibility was determined daily using control precinorm? L in an internal protocol (n=20), within-run CV = 1.3% and between-day CV = 2.013%. The comparison of the HDL-C enzymatic assay with HDL-C precipitating agent was good with a correlation coefficient r = 0.965, Y = 1.981+ 0.959X, n =172. The normal range of HDL-C concentration in serum by this enzymatic method was 36.16 -86.08 mg./dl. Thus, a HDL-C level of over 35 mg/dl., shows the risk is not high and is protective against coronary heart disease.
Description
Keywords
HDL-cholesterol, New enzymatic method
Citation
Tangvarasittichai Surapon, Taytiwat Prawit. Measurement of HDL-cholesterol in serum by a new enzymatic method. Siriraj Medical Journal, 2001 Jul; 53(7): 414-419.