Feasibility study of a novel intraosseous device in adult human cadavers.

dc.contributor.authorSingh, Sandeep
dc.contributor.authorAggarwal, Praveen
dc.contributor.authorLodha, Rakesh
dc.contributor.authorAgarwal, Ramesh
dc.contributor.authorGupta, Arun Kr.
dc.contributor.authorDhingra, Renu
dc.contributor.authorKarve, Jayant Sitaram
dc.contributor.authorJaggu, Srinivas Kiran
dc.contributor.authorBhargava, Balram
dc.date.accessioned2016-08-05T09:03:24Z
dc.date.available2016-08-05T09:03:24Z
dc.date.issued2016-03
dc.description.abstractBackground & objectives: Intraosseous (IO) access is an alternative to difficult intravenous (iv) access during emergency clinical situations. Existing IO solutions are expensive, require power supply and trained manpower; limiting their use in resource constrained settings. To address these limitations, a novel IO device has been developed. The objectives of this study were to evaluate functionality and safety of this device in adult human cadavers. Methods: The ability of the IO device to penetrate the proximal and/or distal tibia was evaluated in three adult cadavers. Subjective parameters of loss of resistance, stable needle hold, easy needle withdrawal and any damage to the device were evaluated during the study. The insertion time was the objective parameter measured. Four sets of radiographs per insertion confirmed the position of the needle and identified complications. Results: A single physician performed 12 IO access procedures using the same device. Penetration of proximal and/or distal tibia was achieved in all instances. It was successful in the first attempt in eight (66.7%) and during second attempt in the remaining. The mean time to insertion was 4.1 ± 3.1 sec. Appropriate insertion of needle in the intra-medullary space of bone was confirmed with radiological examination in 10 (83.3%) insertions. In two occasions after penetrating the cortical layer of bone, the device overshot the intra-medullary space, as detected by radiological examination. Device got bent during insertion in one instance. There was no evidence of needle breakage or bone fracture. The needle could be withdrawn effortlessly in all instances. Interpretation & conclusions: The novel IO device could successfully penetrate the adult cadaver bones in most cases. Further studies are needed to confirm these results on a large sample.en_US
dc.identifier.citationSingh Sandeep, Aggarwal Praveen, Lodha Rakesh, Agarwal Ramesh, Gupta Arun Kr, Dhingra Renu, Karve Jayant Sitaram, Jaggu Srinivas Kiran, Bhargava Balram. Feasibility study of a novel intraosseous device in adult human cadavers. Indian Journal of Medical Research. 2016 Mar; 143(3): 275-280.en_US
dc.identifier.urihttps://imsear.searo.who.int/handle/123456789/176451
dc.language.isoenen_US
dc.source.urihttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC4892072/en_US
dc.subjectDifficult intravenous accessen_US
dc.subjectemergency medicineen_US
dc.subjectintraosseous accessen_US
dc.subjectintraosseous deviceen_US
dc.subjectresuscitationen_US
dc.subjectvascular accessen_US
dc.titleFeasibility study of a novel intraosseous device in adult human cadavers.en_US
dc.typeArticleen_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
ijmr2016v143n3p275.pdf
Size:
1.62 MB
Format:
Adobe Portable Document Format
Description:
Original articles
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: