dc.description.abstractBACKGROUND: Defibrillation testing can be done either via an external cardiac defibrillator or directly via the implanted defibrillator during implantation (device-based testing). The advantage of one testing methodology over the other has not been adequately studied. METHODS AND RESULTS: Seventy-four patients (72% men) were randomized into two groups depending on the defibrillation testing methodology used--external cardiac defibrillation and device-based testing groups. R-wave, pacing threshold, pacing impedance, defibrillation threshold, defibrillation pathway impedance and total procedure time were not significantly different between the two groups. CONCLUSIONS: Device-based testing did not significantly reduce the procedure time. Lead and defibrillation parameters were similar in both the groups; lead repositioning and replacement were required in three patients in the external cardiac defibrillation group.en_US
dc.description.affiliationThoracic and Cardiovascular Institute, Michigan State University and Sparrow Health System, Lansing, MI, USA. anoopgupta@msn.comen_US
dc.identifier.citationGupta AK, Khasnis A, Thakur RK, Lokhandwala Y. Does device-based testing save time during automatic implantable cardioverter-defibrillator implantation? Indian Heart Journal. 2004 Jan-Feb; 56(1): 47-9en_US
dc.identifier.urihttps://imsear.searo.who.int/handle/123456789/3181
dc.language.isoengen_US
dc.source.urihttps://indianheartjournal.comen_US
dc.subject.meshAgeden_US
dc.subject.meshDefibrillators, Implantableen_US
dc.subject.meshFemaleen_US
dc.subject.meshHumansen_US
dc.subject.meshMaleen_US
dc.subject.meshMaterials Testing --methodsen_US
dc.subject.meshMiddle Ageden_US
dc.subject.meshTime Factorsen_US
dc.titleDoes device-based testing save time during automatic implantable cardioverter-defibrillator implantation?en_US
dc.typeClinical Trialen_US
dc.typeJournal Articleen_US
dc.typeRandomized Controlled Trialen_US
Files
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.79 KB
Format:
Plain Text
Description: