A modified formula for defining tissue phantom ratio of photon beams.

dc.contributor.authorAlam, M Jahangiren_US
dc.contributor.authorRabbani, K Sen_US
dc.contributor.authorZakaria, G Aen_US
dc.contributor.authorHussain, S M Akramen_US
dc.contributor.authorKiber, Adnanen_US
dc.contributor.authorBaig, Tanviren_US
dc.date.accessioned2007-12-12en_US
dc.date.accessioned2009-05-27T03:47:03Z
dc.date.available2007-12-12en_US
dc.date.available2009-05-27T03:47:03Z
dc.date.issued2007-12-12en_US
dc.descriptionBangladesh Medical Research Council Bulletin.en_US
dc.description.abstractTissue phantom ratio (TPR), for square fields of various dimensions has been determined at varying depths in water. The dose in water has been measured at a fixed source-to-surface distance (SSD) of 100 cm and reference depth of 5 cm for 6 MV photon beam of Siemens Linear Accelerator Primus 11 in German Cancer Research Center (DKFZ), Heidelberg, Germany. A modified formula has been developed to calculate the TPR value for isocentric treatment. The present article describes the conversion of the measured data values into a comprehensive and consistent data set by the modified formula, that gives the TPR from Percentage Depth Dose (PDD) with depth as a function of field sizes from 10 mm x 10 mm upto 300 mm x 300 mm) and depth (from 0 mm to 300 mm).en_US
dc.description.affiliationDepartment of Physics, University of Dhaka, Dhaka, Bangladesh.en_US
dc.identifier.citationAlam MJ, Rabbani KS, Zakaria GA, Hussain SM, Kiber A, Baig T. A modified formula for defining tissue phantom ratio of photon beams. Bangladesh Medical Research Council Bulletin. 2007 Dec; 33(3): 92-7en_US
dc.identifier.urihttps://imsear.searo.who.int/handle/123456789/258
dc.language.isoengen_US
dc.source.urihttps://www.banglajol.info/index.php/BMRCB/indexen_US
dc.source.urihttps://www.banglajol.info/index.php/BMRCB/article/view/1140/1152en_US
dc.subject.meshAbsorptiometry, Photon --instrumentationen_US
dc.subject.meshHumansen_US
dc.subject.meshModels, Theoreticalen_US
dc.subject.meshPhantoms, Imagingen_US
dc.subject.meshPhotonsen_US
dc.subject.meshRadiometry --instrumentationen_US
dc.titleA modified formula for defining tissue phantom ratio of photon beams.en_US
dc.typeJournal Articleen_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
bmrcb2007v33n3p92.pdf
Size:
423.4 KB
Format:
Adobe Portable Document Format
Description:
Journal article
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.79 KB
Format:
Plain Text
Description: