Iridociliary measurements using the Anterion swept-source optical coherence tomography

dc.contributor.authorPatyal, Sagarikaen_US
dc.contributor.authorYadav, Arun Kumaren_US
dc.contributor.authorAnthony, Hanselen_US
dc.contributor.authorAnsari, Faisalen_US
dc.date.accessioned2023-08-25T06:35:46Z
dc.date.available2023-08-25T06:35:46Z
dc.date.issued2023-01
dc.description.abstractPurpose: Anterior segment optical coherence tomography (AS?OCT) by a swept source can visualize the ciliary body. The study was performed for analyzing the feasibility of a new swept?source OCT (Anterion) device for measuring iris ciliary sulcus, ciliary body thickness, and iris thickness. Methods: It is a pilot study among 30 normal young participants and hospital employees with their consent. The ciliary body, iris thickness, and ciliary sulcus were measured and compared to the results in the literature. We obtained 28 good?quality results. The main outcome measures were iris thickness (near and away from the pupillary margin), ciliary body thickness, and iridociliary sulcus measurement. Results: The iris thickness 2.5 mm from the pupillary margin was 0.70 ± 0.10 mm in the right eye and 0.68 ± 0.11 in the left eye. Closer to the root, the iris thickness varied from 0.55 ± 0.16 mm (right eye) and 0.57 ± 0.12 (left eye). Ciliary body thickness: Right eye 0.59 ± 0.14mm (28 good images), left eye 0.58 ± 0.13mm; ciliary sulcus (inner angle): right eye 82.65 ± 26.6°, left eye 83.66 ± 30.1°. Conclusion: Anterion OCT is feasible in our setting for various measurements and can provide valuable inputs to ophthalmologists for the treatment of ocular diseases.en_US
dc.identifier.affiliationsCentre for Sight, New Delhi, Indiaen_US
dc.identifier.affiliationsDepartment of Community Medicine, AFMC, Pune, Maharashtra, Indiaen_US
dc.identifier.affiliationsApplication Specialist, Toshbro Medical Pvt Ltd., Delhi, Indiaen_US
dc.identifier.citationPatyal Sagarika, Yadav Arun Kumar, Anthony Hansel, Ansari Faisal. Iridociliary measurements using the Anterion swept-source optical coherence tomography. Indian Journal of Ophthalmology. 2023 Jan; 71(1): 109-112en_US
dc.identifier.issn1998-3689
dc.identifier.issn0301-4738
dc.identifier.placeIndiaen_US
dc.identifier.urihttps://imsear.searo.who.int/handle/123456789/224777
dc.languageenen_US
dc.publisherAll India Ophthalmological Societyen_US
dc.relation.issuenumber1en_US
dc.relation.volume71en_US
dc.source.urihttps://doi.org/10.4103/ijo.IJO_1779_22en_US
dc.subjectCiliary body thicknessen_US
dc.subjectIridociliary sulcusen_US
dc.subjectiris thicknessen_US
dc.subjectswept?source optical coherence deviceen_US
dc.titleIridociliary measurements using the Anterion swept-source optical coherence tomographyen_US
dc.typeJournal Articleen_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
IJO2023v71n1p109.pdf
Size:
1.16 MB
Format:
Adobe Portable Document Format