Impact of external haematology proficiency testing programme on quality of laboratories.
No Thumbnail Available
Date
2007-11-28
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
BACKGROUND & OBJECTIVE: A reliable and reproducible report from a laboratory needs internal quality control within the laboratory and participation in external proficiency testing programmes (EPTP). This study conducted at the Department of Haematology, All India Institute of Medical Sciences (AIIMS), New Delhi, which has been conducting an EPTP since 1992, was undertaken to assess the efficacy of this programme in improving the performance of participating laboratories in reporting test samples sent for Hb, total leucocyte count (TLC), reticulocyte count and assessment of peripheral blood smear (PBS).
METHODS: The samples were prepared in our laboratory according to the International Standards Organization (ISO) guidelines. The performance of individual laboratories was assessed using robust Z score, which is an indicator of acceptability of the test result.
RESULTS: An improvement in the overall percentage of laboratories with acceptable reports was seen during the study period. It has increased from 38,40,40 per cent in 1992 to 85, 90,94.7 per cent in 2006 for Hb, TLC, reticulocyte count, respectively. However, the results for peripheral smear assessment improved only marginally. INTERPRETATION & CONCLUSION: The external haematology proficiency testing programme run by our department for Hb, TLC, reticulocyte count, and peripheral blood smear assessment, has helped in improving the reporting standards of these parameters in Indian laboratories.
Description
Keywords
Citation
Saxena R, Katoch SC, Srinivas U, Rao S, Anand H. Impact of external haematology proficiency testing programme on quality of laboratories. Indian Journal of Medical Research. 2007 Nov; 126(5): 428-32