Is Estimation Of Blood Glucose By Glucometer More Reliable Than Laboratory Analysis? A Comparative Study Done At Rural Hospital Of Maharashtra

dc.contributor.authorParveen, Shaistaen_US
dc.contributor.authorKausar, Afshanen_US
dc.date.accessioned2023-07-14T07:52:24Z
dc.date.available2023-07-14T07:52:24Z
dc.date.issued2022-08
dc.description.abstractIntroduction: Diabetes is a major public health problem that is approaching epidemic propor?ons globally. Selfmonitoring of blood glucose (SMBG) by the American Diabetes Associa?on (ADA) is promoted in diabe?c pa?ents so that they achieve and maintain specific glycemic goals. Objec?ves : To compare the blood glucose es?ma?on methods; capillary blood and venous blood by glucometer, venous plasma glucose es?ma?on by auto analyzer and to find varia?on in results. Material & Methods : 60 pa?ents a?ending Outpa?ent department of JIIUs IIMSR and Noor hospital, Warudi, Badnapur, Maharashtra who were advised blood glucose es?ma?on were selected. Finger prick (capillary) blood glucose & glucose es?ma?on of venous blood was done by glucometer; and venous plasma glucose es?ma?on was done by auto analyzer in laboratory. Result & Conclusion : Mean values of capillary glucose es?ma?on by glucometer were higher than plasma glucose es?ma?on in laboratory on auto analyzer and the difference was found to be sta?s?cally significant. Venous glucose es?ma?on on glucometer gave high erroneous results as compared to plasma glucose es?ma?on in laboratory.en_US
dc.identifier.affiliationsAssociate Professor ,Department of Physiology, JIIUs IIMSR Medical College, Dept. of Physiology, JIIUs IIMSR Medical College, Warudi, Maharashtra, Indiaen_US
dc.identifier.citationParveen Shaista, Kausar Afshan. Is Estimation Of Blood Glucose By Glucometer More Reliable Than Laboratory Analysis? A Comparative Study Done At Rural Hospital Of Maharashtra. Prathima Institute of Medical Sciences. 2022 Aug; 10(2): 9-12en_US
dc.identifier.issn2348-229X
dc.identifier.issn2348-1447
dc.identifier.placeIndiaen_US
dc.identifier.urihttps://imsear.searo.who.int/handle/123456789/218973
dc.languageenen_US
dc.publisherPrathima Institute of Medical Sciencesen_US
dc.relation.issuenumber2en_US
dc.relation.volume10en_US
dc.source.urihttps://doi.org/10.47799/pimr.1002.03en_US
dc.subjectAutoanalyzeren_US
dc.subjectBloodglucoseen_US
dc.subjectGlucometeren_US
dc.subjectSelfmonitoring of blood glucoseen_US
dc.titleIs Estimation Of Blood Glucose By Glucometer More Reliable Than Laboratory Analysis? A Comparative Study Done At Rural Hospital Of Maharashtraen_US