New method of electrode placement for determination of cardiac output using impedance cardiography.

dc.contributor.authorBarde, Pradip Ben_US
dc.contributor.authorJindal, Ghanshyam Den_US
dc.contributor.authorSingh, Rajviren_US
dc.contributor.authorDeepak, Kishore Ken_US
dc.date.accessioned2006-07-30en_US
dc.date.accessioned2009-06-01T07:28:44Z
dc.date.available2006-07-30en_US
dc.date.available2009-06-01T07:28:44Z
dc.date.issued2006-07-30en_US
dc.description.abstractIn thirty eight young healthy adult subjects, cardiac output was determined non-invasively by using two methods of electrode placement viz. vertical (uses silver braided wires in a band shape) and horizontal (conventional stick-on type surface ECG electrodes), using the technique of Impedance Cardiography. The recordings were taken in supine position on the same day in two separate sets with 30.0 min interval between two sets. In each set of recording, five successive recordings, each at an interval of five minutes were taken. The mean values of cardiac out put by two methods were compared. For the set I, the respective values (Mean +/- SD) of cardiac output by horizontal and vertical methods for the each of the five recordings were 4.87 +/- 0.77 and 5.03 +/- 0.64 for the first, 4.87 +/- 0.71 and 4.91 +/- 0.66 for the second, 4.99 +/- 0.67 and 5.00 +/- 0.70 for the third, 4.78 +/- 0.69 and 4.98 +/- 0.61 for the fourth, 4.84 +/- 0.69 and 4.98 +/- 0.62 for the fifth recording in supine position. The respective P values for these pairs for between the group comparisons were 0.33, 0.50, 0.96, 0.17, and 0.36. In addition, to see the repeatability for each method, within the group comparison was done, the P values were 0.71 and 0.91 for the horizontal and vertical methods, respectively. The mean value of cardiac output did not differ significantly between two methods for recordings of set II in supine position. The cardiac output measurement by placing four spot electrodes horizontally, gave consistent result on repeated measurements and their values showed concordance with the cardiac output values obtained by conventional four band electrodes tied around the chest.en_US
dc.description.affiliationDepartment of Physiology, All India Institute of Medical Sciences, New Delhi 110 029.en_US
dc.identifier.citationBarde PB, Jindal GD, Singh R, Deepak KK. New method of electrode placement for determination of cardiac output using impedance cardiography. Indian Journal of Physiology and Pharmacology. 2006 Jul-Sep; 50(3): 234-40en_US
dc.identifier.urihttps://imsear.searo.who.int/handle/123456789/108185
dc.language.isoengen_US
dc.source.urihttps://www.ijpp.comen_US
dc.subject.meshAdulten_US
dc.subject.meshCardiac Output --physiologyen_US
dc.subject.meshCardiography, Impedance --methodsen_US
dc.subject.meshElectrodesen_US
dc.subject.meshHumansen_US
dc.subject.meshMaleen_US
dc.titleNew method of electrode placement for determination of cardiac output using impedance cardiography.en_US
dc.typeComparative Studyen_US
dc.typeJournal Articleen_US