Pressure bandage over venous conduit harvesting site causing compartment syndrome in a patient with intra-aortic balloon pump: An unusual cause.
dc.contributor.author | Rastogi, Amit | |
dc.contributor.author | Haldar, Rudrashish | |
dc.contributor.author | Majumdar, Gauranga | |
dc.contributor.author | Pahade, Akhilesh, | |
dc.contributor.author | Singh, Prabhat K | |
dc.date.accessioned | 2015-08-25T04:35:21Z | |
dc.date.available | 2015-08-25T04:35:21Z | |
dc.date.issued | 2015-07 | |
dc.description.abstract | Intra Aortic Balloon Pump (IABP) is conventionally used to support coronary perfusion and weaning from cardiopulmonary bypass. IABP in situ has its own share of complications. We present a case where a patient on IABP support who had reduced peripheral pulsations of the ipsilateral limb and was initially misdiagnosed as IABP catheter associated thromboembolism. A negative embolectomy ruled out the same. Further looking for the cause of reduction of ipsilateral pulses it was found that the tight compressive bandage at saphenous vein conduit harvesting site had led to development of compartment syndrome (CS). | en_US |
dc.identifier.citation | Rastogi Amit, Haldar Rudrashish, Majumdar Gauranga, Pahade Akhilesh, Singh Prabhat. K. Pressure bandage over venous conduit harvesting site causing compartment syndrome in a patient with intra-aortic balloon pump: An unusual cause. Annals of Cardiac Anaesthesia. 2015 Jul; 18(3): 453-459. | en_US |
dc.identifier.uri | https://imsear.searo.who.int/handle/123456789/162402 | |
dc.language.iso | en | en_US |
dc.source.uri | https://www.annals.in/article.asp?issn=0971-9784;year=2015;volume=18;issue=3;spage=453;epage=455;aulast=Rastogi | en_US |
dc.subject | Compartment syndrome | en_US |
dc.subject | Thromboembolism | en_US |
dc.subject | Venous conduit | en_US |
dc.subject.mesh | Adult | |
dc.subject.mesh | Blood Vessel Prosthesis | |
dc.subject.mesh | Compartment Syndromes --epidemiology | |
dc.subject.mesh | Compartment Syndromes --etiology | |
dc.subject.mesh | Compression Bandages --adverse effects | |
dc.subject.mesh | Humans | |
dc.subject.mesh | Intra-Aortic Balloon Pumping --epidemiology | |
dc.subject.mesh | Male | |
dc.subject.mesh | Pressure | |
dc.subject.mesh | Saphenous Vein | |
dc.subject.mesh | Thromboembolism --epidemiology | |
dc.subject.mesh | Thromboembolism --etiology | |
dc.subject.mesh | Tissue and Organ Harvesting | |
dc.title | Pressure bandage over venous conduit harvesting site causing compartment syndrome in a patient with intra-aortic balloon pump: An unusual cause. | en_US |
dc.type | Article | en_US |