Descemet's tear due to injector cartridge tip deformity.

dc.contributor.authorBiswas, Partha
dc.contributor.authorSengupta, Subhrangshu
dc.contributor.authorPaul, Ajoy
dc.contributor.authorKochgaway, Lav
dc.contributor.authorBiswas, Sayantan
dc.date.accessioned2012-07-12T09:16:23Z
dc.date.available2012-07-12T09:16:23Z
dc.date.issued2012-05
dc.description.abstractFoldable intraocular lens (IOL) implantation using an injector system through 2.8-mm clear corneal incision following phacoemulsification provides excellent speedy postoperative recovery. In our reported case, a Sensar AR40e IOL (Abbott Medical Optics, USA) was loaded into Emerald C cartridge, outside the view of the operating microscope, by the first assistant. The surgeon proceeded with the IOL injection through a 2.8-mm clear corneal incision after uneventful phacoemulsification, immediately following which he noted a Descemet's tear with a rolled out flap of about 2 mm near the incision site. Gross downward beaking of the bevelled anterior end of the cartridge was subsequently noticed upon examination under the microscope. We suggest careful preoperative microscopic inspection of all instruments and devices entering the patient's eyes to ensure maximum safety to the patient.en_US
dc.identifier.citationBiswas Partha, Sengupta Subhrangshu, Paul Ajoy, Kochgaway Lav, Biswas Sayantan. Descemet's tear due to injector cartridge tip deformity. Indian Journal of Ophthalmology. 2012 May; 60(3): 218-220.en_US
dc.identifier.urihttps://imsear.searo.who.int/handle/123456789/139475
dc.language.isoenen_US
dc.source.urihttps://www.ncbi.nlm.nih.gov/pmc/articles/PMC3361820/en_US
dc.subjectCartridge tipen_US
dc.subjectdescemet's tearen_US
dc.subjectintraocular Lensen_US
dc.subjectphacoemulsificationen_US
dc.subject.meshCartilage
dc.subject.meshDescemet Membrane --injuries
dc.subject.meshEquipment Failure
dc.subject.meshEye Injuries --etiology
dc.subject.meshHumans
dc.subject.meshInjections --adverse effects
dc.subject.meshInjections --instrumentation
dc.subject.meshIntraoperative Complications
dc.subject.meshLens Implantation, Intraocular --adverse effects
dc.subject.meshLens Implantation, Intraocular --instrumentation
dc.subject.meshRupture
dc.titleDescemet's tear due to injector cartridge tip deformity.en_US
dc.typeArticleen_US
Files
Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
ijo2012v60n3p218.pdf
Size:
836.66 KB
Format:
Adobe Portable Document Format
Description:
Journal article
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: