Validation of ESR analyzer using Westergren ESR method.

No Thumbnail Available
Date
2007-07-22
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Erythrocyte sedimentation rate (ESR) is one of the most frequently ordered laboratory test. ESR analyzers were developed to provide a quick and efficient measure of ESR. We compared the results of ESR obtained by an ESR analyzer with those by the Westergren method in a group of 75 patients Linear regression analysis showed a good correlation between the two results (r = 0.818, p < 0.01). The intra class correlation was 0.82. The analyzer method had the advantages of safety, decreased technician time and improved patient care by providing quick results.
Description
Keywords
Citation
Sikka M, Tandon R, Rusia U, Madan N. Validation of ESR analyzer using Westergren ESR method. Indian Journal of Pathology & Microbiology. 2007 Jul; 50(3): 634-5