Sterilisation of tonometers and gonioscopes.

No Thumbnail Available
Date
1998-06-16
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Precautions to prevent spread of infection through tonometers and gonioscopes are described in this article. Tonometers and gonioscopes should not be used in the presence of clinically manifest conjunctivitis and keratitis. The Schiotz tonometer should be dipped in a 1:1000 merthiolate solution, and rinsed in saline/distilled water prior to use. The Goldmann applanation prism tip can be wiped with gauze soaked in 70% isopropyl alcohol and then dried before use. Gonioscopes should be cleaned in running water, wiped with gauze soaked in 70% isopropyl alcohol, and then dried before use. Koeppes and goniotomy lenses can be sterilized with ethylene oxide, prior to use in surgery.
Description
15 references.
Keywords
Citation
Sood D, Honavar SG. Sterilisation of tonometers and gonioscopes. Indian Journal of Ophthalmology. 1998 Jun; 46(2): 113-6