Ultrasonic errors in analysis of vitreous haemorrhage.

No Thumbnail Available
Date
1990-10-01
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
Pre-vitrectomy errors in diagnostic A & B mode ultrasound may result in affection of the surgical outcome. We have done a retrospective analysis of fifty patients who were screened by both A & B mode echography and analyzed the data which revealed that real-time kinetic echography is required for accurate interpretation of the vitreoretinal status. Besides, quantitative A scan examination also has fallacies due to the echo beam not striking the sclerochoroid complex at right angles, and a modification of this technique is suggested. Lastly old retinal detachments may not produce echo-patterns diagnostic of retinal detachments resulting in fallacious interpretation. We recommend that combined A mode and real time B-mode and often repeat echography is essential for accuracy in diagnosis.
Description
Keywords
Citation
Kumar A, Verma L, Jha SN, Tewari HK, Khosla PK. Ultrasonic errors in analysis of vitreous haemorrhage. Indian Journal of Ophthalmology. 1990 Oct-Dec; 38(4): 162-3