Can ocular torsion be measured using the slitlamp biomicroscope?

No Thumbnail Available
Date
2005-03-15
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
PURPOSE: To evaluate and compare a stereo-biomicroscopic technique of ocular torsion measurement with fundus photographic technique. METHODS: This prospective masked observational study included 72 eyes of 36 consecutive subjects with normal ocular motility. The disc-foveal angle was measured by two masked investigators using a slitlamp stereo-biomicroscopic and fundus photographic technique. RESULTS: The mean age was 13.7 years (6 - 44 years), and 15 (41.6%) were males. The average torsion using the slitlamp technique was 5.5 +/- 3.3 degrees and 6.1 +/- 4.3 degrees with fundus photographic technique. Clinical agreement between the two techniques for 5 degree and 6 degree torsion was 81% and 90% respectively. The average difference between the two techniques was 3.0 +/- 2.6 degrees . The mean inter-ocular difference was 3.3 +/- 2.7 degrees (SD, 2.77) and 5.5 +/- 4.6 degrees using the slitlamp and fundus photographic technique respectively. The Pearson correlation coefficient (r2) was 0.5. CONCLUSION: The fundus photographic technique remains the gold standard test for the measurement of ocular torsion. The slitlamp biomicroscopic method can be useful where a fundus camera is unavailable. Further studies are required to test the repeatability of these tests.
Description
Keywords
Citation
Kothari MT, Venkatesan G, Shah JP, Kothari K, Nirmalan PK. Can ocular torsion be measured using the slitlamp biomicroscope? Indian Journal of Ophthalmology. 2005 Mar; 53(1): 43-7