Acetabular position setting in total hip arthoplasty by using V-inclinometer.

No Thumbnail Available
Date
2004-04-26
Journal Title
Journal ISSN
Volume Title
Publisher
Abstract
The abduction angle of the acetabular implant position is important in the process of total hip arthroplasty. The authors evaluated the potential benefit of a V-inclinometer in the prediction of the abduction angle. To evaluate the accuracy of the V-inclinometer, acetabular cups were inserted in 50 cadaveric pelvises. The abduction angle from V-inclinometer and digital photograph were compared and analyzed using paired t-test. There was no statistical difference of the acetabular abduction angle between digital photograph and V-inclinometer (P > 0.05). The V-inclinometer can be used to predict the abduction angle on a cadaveric pelvis, suggesting that it could be a useful adjunct in clinical practice.
Description
Chotmaihet Thangphaet.
Keywords
Citation
Wilairatana V, Prasongchin P. Acetabular position setting in total hip arthoplasty by using V-inclinometer. Journal of the Medical Association of Thailand. 2004 Apr; 87(4): 353-6