Evaluation of efficacy of BIORAD D10TM testing system in detection of beta thalassaemia carrier

Loading...
Thumbnail Image
Date
2019-03
Journal Title
Journal ISSN
Volume Title
Publisher
Medip Academy
Abstract
Background: Hemoglobinopathies especially thalassaemia and its interaction with HbE and HbS are significant cause of morbidity and mortality in our country. With no feasible treatment, prevention of cases by carrier detection is the only option for successful reduction of the disease burden. VARIANT hemoglobin testing system BIORAD using principle of cation exchange chromatography has been in use and considered as gold standard for carrier detection. The purpose of this study is to compare the efficacy of a different, cheaper instrument; D10 by the same manufacturer BIORAD for carrier detection in beta thalassaemia.Methods: Patients diagnosed as beta thalassaemia carrier by VARIANT hemoglobin testing system (HbA2 value between 4.0-9.0) were retested using D10 instrument and checked for agreement.Results: There was good correlation between VARIANT and D-10 methods with Intraclass correlation coefficient 0.832 (95% Confidence Interval 0.756-0.884). Bland-Altman analysis showed mean bias of +0.3526 (95% CI -0.3958 to +1.101).Conclusions: Although further study is needed with larger sample size for assessment of sensitivity and specificity of D10 instrument, it is evident from this study that this instrument can be an effective and cheaper alternative of VARIANT hemoglobin testing system.
Description
Keywords
Beta thalassaemia, BIORAD D10, BIORAD variant, Carrier, Detection
Citation
Roy Debajyoti Singha, Bhattacharyya Riju, Mukhopadhyay Kaushik, Bandopadhyay Debasis. Evaluation of efficacy of BIORAD D10TM testing system in detection of beta thalassaemia carrier. International Journal of Research in Medical Sciences. 2019 Mar; 7(3): 782-785