Abstract
Fluctuations in core body temperature beyond a narrow range are of concern as they indicate
changed homeostasis. Thermoregulation remains a key aspect of neonatal care. New ‘non
touch’ method of measuring temperature may reduce infection rate and discomfort of
neonates. Body temperature of neonates admitted in Neonatal Intensive Care Unit was
measured using axillary digital thermometer and a handheld infrared non touch thermometer.
The two methods did not agree well (mean difference = -0.5, 95% limits of agreement: [-2.3,
1.2]). The agreement was similar with a negligible difference when patients in open care
warmers were excluded (mean difference = -0.6, 95% limits of agreement: [-2.3, 1.1]). As the
accuracy is unsatisfactory, the said technology needs further improvement before it can be
adapted for routine patient care.